The WALOP-North Instrument I: Optical Design, Filter Design, Calibration

1 Dec 2024arXiv:2412.00964links table onlyarchive 2025-07-28

John A. Kypriotakis, Siddharth Maharana, Ramya M. Anche, Chaitanya V. Rajarshi, A. N. Ramaprakash, Bhushan Joshi, Artem Basyrov, Dmitry Blinov, Tuhin Ghosh, Eirik Gjerlow, Sebastian Kiehlmann, Nikolaos Mandarakas, Georgia V. Panopoulou, Katerina Papadaki, Vasiliki Pavlidou, Timothy J. Pearson, Vincent Pelgrims, Stephen B. Potter, Anthony C. S. Readhead, Raphael Skalidis, Konstantinos Tassis

The archive published only this paper's code-link row. Authors, date and abstract are from arXiv's metadata (CC0), read from the Kaggle arXiv metadata snapshot of 2026-09-12 where its title matched the archive's; the title is the archive's.

The Wide Area Linear Optical Polarimeter North (WALOP-North) is an optical polarimeter designed for the needs of the PASIPHAE survey. It will be installed on the 1.3m telescope at the Skinakas Observatory in Crete, Greece. After commissioning, it will measure the polarization of millions of stars at high Galactic latitude, aiming to measure hundreds of stars per deg². The astronomical filter used in the instrument is a modified, polarimetrically-neutral broadband SDSS-r. This instrument will be pioneering one due to its large field-of-view (FoV) of 30×30 arcmin² and high accuracy polarimetry measurements. The accuracy and sensitivity of the instrument in polarization fraction will be at the 0.1\% and 0.05\% level, respectively. Four separate 4k×4k CCDs will be used as the instrument detectors, each imaging one of the 0, 45, 90 and 135 polarized FoV separately, therefore making the instrument a four-channel, one-shot polarimeter. Here, we present the overall optical design of the instrument, emphasizing on the aspects of the instrument that are different from WALOP-South. We also present a novel design of filters appropriate for polarimetry along with details on the management of the instrument size and its polarimetric calibration.

PaperPDFCode

Code

Repository list and official/mentioned flags are the archive's, frozen 2025-07-28. Reachability, where shown, is from one Syntology probe window (2026-09-16 to 2026-09-18); repositories not probed show nothing. GitHub stars are not tracked.

Code Syntology ran Syntology

Not run by Syntology. Nothing on this page verifies that the listed code works.

Results from the paper archive 2025-07-28

No leaderboard rows for this paper in the archive.

Report a problem or propose a change · a person checks every report against the paper or source before anything changes; decisions are listed on /corrections